ICANL
Records of service and maintenance must
be maintained.
Imaging equipment parameters include:
Energy
peaking - daily
Intrinsic
or extrinsic uniformity - daily
Resolution
and linearity - weekly
High
count calibration floods - monthly
Center
of rotation – monthly
Collimator
integrity – annually
Preventive
maintenance - 6 months, or per
manufacturer’s recommendations
PET Quality Control
Blank
scan - daily
Normalization
– After hardware change or
per manufacturer specification
Absolute
activity calibration for SUV – After hardware change or per manufacturer specification.
Preventive
maintenance - every 6 months
Non-Imaging Equipment Quality Control
3.3.1- A policy must exist and be followed for routine inspection and testing of all non-imaging equipment such as dose calibrators, uptake probes, survey meters and glucometers and be in accordance with the federal, state and local requirements. The dose calibrator must be calibrated in accordance with nationally recognized standards or the manufacturers instructions.
| Equipment | Test | Frequency |
| Dose Calibrator | Constancy | Daily |
| Dose Calibrator | Linearity | Quarterly |
| Dose Calibrator | Accuracy | Annually |
| Survey Meter | Calibration | Annually |
| Glucometer | Accuracy | Daily |