PET Quality Control

Non-Imaging Equipment Quality Control

3.3.1- A policy must exist and be followed for routine inspection and testing of all non-imaging equipment such as dose calibrators, uptake probes, survey meters and glucometers and be in accordance with the federal, state and local requirements. The dose calibrator must be calibrated in accordance with nationally recognized standards or the manufacturers instructions.

Equipment Test Frequency
Dose Calibrator Constancy Daily
Dose Calibrator Linearity Quarterly
Dose Calibrator Accuracy Annually
Survey Meter Calibration Annually
Glucometer Accuracy Daily